Abstract

The electrical motion of the heart is characterised by the ECG signal.
ECG elucidation can be used to detect the heart syndrome. This technology has
an efficient diagnostic tool, due to the high regard of portable electronic
products, low power system has fascinated more consideration in recent years.
This work presents digital ECG data acquisition system to diminish the power
consumption. In the proposed work, analogue block is not used, they convert
the input voltage into a digital code by delay lines. This digital architecture is
capable of operating with a low supply voltage such as 0.3 V and 0.1 V. In this
architecture, analogue blocks such as low-noise amplifier (LNA) and filters
are not used. A digital feedback loop is employed to cancel the impact of the
dc offset on the circuit, which eliminates the requirements for coupling
capacitors. The circuit is implemented in 130 nm and 65 nm CMOS process.
The simulation results illustrate that the front-end circuit of digital architecture
for 130 nm consumes 18.9 pW and 65 nm consumes 109 pW of power.

Item Type: Article
Subjects: Biomedical Engineering > Biomedical Instrumentation
Divisions: Biomedical Engineering
Depositing User: Unnamed user with email techsupport@mosys.org
Date Deposited: 30 Oct 2026 04:56
Last Modified: 30 Oct 2026 04:56
URI: https://ir.vcenggw.ac.in/id/eprint/529

Actions (login required)

View Item
View Item